Changes between Initial Version and Version 1 of GPC1_BrightStarAstrometry_SW_LH_scan


Ignore:
Timestamp:
Apr 20, 2011, 6:45:22 PM (15 years ago)
Author:
watersc1
Comment:

--

Legend:

Unmodified
Added
Removed
Modified
  • GPC1_BrightStarAstrometry_SW_LH_scan

    v1 v1  
     1= Second Microtest Set =
     2
     3This data set consisted of two exposures taken with the default voltages, one with the microtest 05 voltages (SW_L = default - 1.0 V, SW_H = default - 1.0 V), and six others that vary only SW_L or SW_H (with the other set to the default value).  To study the impact of these changes, I reduced the data as before, and found the best fit lines in the dx and dy offsets as a function of instrumental magnitudes.  The data were binned using 0.1 magnitude bins, and the fit performed between -14 and -12 (other fits were performed using ranges (-16:-14) and (-12:-10), but are not shown here).
     4
     5The following images show the best fit slopes as a function of voltage in both dx and dy directions, with the dy data providing a control.  These show convincingly that SW_L has little effect on the slope within this range, but SW_H values below -0.5 decrease the slope significantly.  Comparing the 2-phase and 3-phase devices show that the 3-phase devices have generally acceptable slopes regardless of SW_L and SW_H.
     6
     7|| OTA33 / 3-phase              || OTA34 / 2-phase               ||
     8|| [[Image(SW_tests_OTA33.png)]]|| [[Image(SW_tests_OTA34.png)]] ||
     9
     10[[Image(SW_tests_montage_big.png,1024px)]]
     11
     12
     13 * [wiki:GPC_BrightStarAstrometry_Microtest Return]